Multistage Roots vacuum pumps are a type of positive displacement pumps that utilize two or more intermeshing rotors to compress gas. They are commonly used in the analytical instrument applications for their high pumping speed, low ultimate pressure, and clean pumping environment. They can achieve high vacuum levels at lowest noise levels without the need for oil, leading to low maintenance and high reliability.
Introducing the nXLi and nXRi pumps from Edwards, designed for high-end mass spectrometry applications. The nXLi pumps offer enhanced performance and reliability, while the nXRi pumps feature a compact design, perfect for space-saving integration into analytical instruments and vacuum systems.
Explore our small dry pumps collection for efficient, oil-free vacuuming solutions tailored to your specific needs!
Edwards is delighted to introduce the new air cooled nXLi dry pump range, designed to deliver a consistent performance worldwide with complete user flexibility in mind. This new single phase pump is available in two options: nXL110i and nXL200i. Both are optimised for LCMS and ICPMS duties and are capable of handling gas loads of up to 25 slm. nXLi can also be used on chambers up to 100 litres, at continuous inlet pressures up to 15 mbar with occasional pump down.
General Product Information
Connection Inlet Flange
DN 40 ISO-KF (NW40)
Weight
84.00 kg
Connection vacuum outlet flange
DN 25 ISO-KF (NW25)
Leak Rate
1.000 × 10?? pa·m3/s
Power Supply Phase
1-Ph
Power Supply Voltage
1Ph 200-240V 50/60Hz
Dimension Length
0.704 m
IP Rating
IP20
Power Supply Frequency
50 Hz
60 Hz
Dimension Height
0.343 m
Dimension Width
0.308 m
Connector Electrical Power
IEC EN60320 C20 - Male (for C19 plug on mains cable)
Noise / Sound Level
55.40 db(A)
Cooling Method
Air Cooling (forced - fan)
Vibration Level Velocity (rms)
2.500 × 10?3 m/s
Connection Inlet / Outlet Flange
DN 40 ISO-KF (NW40) / DN 25 ISO-KF (NW25)
Accessories / Scope of Supply
This product does not include the correct local mains power cable. Has to be additionally selected from the list of suitable accessories.
Dimension Comments
The dimensions length, width and height try to describe the outer shape of this particular product in its typical orientation when standing on the ground. We always recommend checking the detailed dimensions given in 2D drawings in the instruction manual or the 3D shapes given in 3D models.
Certification Motor
UL
CSA
Performance Data
Power Total Rated
700.00 W
Pressure Ultimate
2.00 Pa
Nominal rotational speed
150.00 Hz
Power at Ultimate
650.00 W
Pumping Speed Air Nitrogen (@50Hz or max @default setting)
0.056 m3/s
Operation Conditions - Limitations
Pressure Absolute Inlet Continous Max
800.00 Pa
Altitude Max
2000.00 m
Pressure Absolute Outlet Continous Max
1.500 × 10? Pa
Environmental Conditions
Temperature Ambient Operation Max
313.15 K
Temperature Ambient Operation Min
278.15 K
Temperature Ambient Storage Min
243.15 K
Temperature Ambient Storage Max
343.15 K
Classification
Classification Product Pressure Range
Medium Vacuum: 1 mbar - 1·10?3 mbar
Classification High / Medium / Rough Vacuum pump
Rough Vacuum Pump
Classification Dry or Wet Pump
Dry
Classification Vacuum Pump Technology
Multistage Roots Pump
Material of Construction
Material of Construction
SG iron,Aluminium alloy,Steel,Fluoroelastomer (seals),PTFE
Connectivity - Communication Interface
Connection Accessories Connector
Phoenix Contact M8 4-pin connector Male (SACC-DSI-M 8FS-4CON-M12/0.5)